Article ID Journal Published Year Pages File Type
1215619 Journal of Chromatography B 2007 8 Pages PDF
Abstract
A LC-MS/MS method was validated for the determination of BA011FZ041, a styrylquinoline derivative. After addition of BA011FZ055 as internal standard (IS), the method involved solid phase extraction (SPE), LC separation with an ether-phenyl column and quantification by MS/MS after positive ESI. The calibration curve, ranging from 1 to 500 ng/mL was fitted to a 1/x-weighted quadratic regression model. Lower limit of quantification (LLOQ) was 1 ng/mL using 100 μL of plasma. Intra- and inter-assay precision and accuracy values were within the regulatory limits. The method was successfully applied to the determination of BA011FZ041 in rat plasma and PBMCs after i.v. dosing.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,