Article ID Journal Published Year Pages File Type
1215967 Journal of Chromatography B 2007 6 Pages PDF
Abstract

A rapid, reproducible and accurate high-performance liquid chromatographic (HPLC) method for the quantitative determination of sphingomyelin in rat brain was developed and validated using normal-phase silica gel column, acetonitrile–methanol–water (65:18:17 (v/v)) at a flow rate of 1 ml/min, isocratic elution, UV detection at 207 nm and 1,2-dimyristoyl-sn-glycero-3-phosphocholine as an internal standard. Total run time was 10.0 min. The calibration curve was linear over the range of 0.025–0.4 mg/ml sphingomyelin (R2 > 0.99). The intra-day coefficient of variation ranged from 1.4% to 2.2%. The average inter-day coefficient of variation over a period of 4 days was 3.1%. The practical limit of detection was 0.005 mg/ml with a quantification limit of 0.01 mg/ml.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,