Article ID Journal Published Year Pages File Type
1216157 Journal of Chromatography B 2006 5 Pages PDF
Abstract

In this paper the determination of Ochratoxin A (OTA) in low volumes of human blood serum by enzyme-linked immunosorbent assay (ELISA) is compared with an appropriate capillary electrophoresis with laser-induced fluorescence detection (CE-LIF) method. In order to use ELISA for high-throughput analysis in epidemiological studies no sample cleanup was performed. Both methods showed a limit of detection (LOD) of 0.5 ng/mL. Comparing the precisions of both methods, the data show that the quantified concentrations in ELISA are higher than the corresponding concentrations in the CE-LIF method. Using a matrix calibration curve instead of a standard calibration curve the reproducibilities of both methods are comparable. No additional matrix effect could be observed by adding phenylalanine as probable matrix compound to the serum.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry