Article ID Journal Published Year Pages File Type
1216511 Journal of Chromatography B 2006 7 Pages PDF
Abstract

Quaternary ammonium drugs (atracurium, bretylium, edrophonium, ipratropium, mivacurium, neostigmine, pancuronium and rocuronium) and herbicides (difenzoquat, diquat and paraquat) in human whole blood were analysed by LC/MS/MS with positive electrospray ionisation (ESI), following extraction with Bond Elut® LRC-CBA cartridges. Internal standards were benzyldimethylphenylammonium chloride monohydrate and ethyl viologen for drug and herbicide analysis, respectively. Ion-pair chromatography used heptafluorobutyric acid (15 mM)–ammonium formate (20 mM) buffer adjusted to pH 3.30 with formic acid and a linear gradient from 5 to 90% methanol run over 18 min. Recoveries ranged from 79.7 to 105.1%, detection limits were between 3.6 and 20.4 ng/ml and the intra- and inter-day precisions were less than 18.6% at a concentration of 10 ng/ml. The method was applied to a case of accidental paraquat poisoning in which the concentration of paraquat in blood was 0.64 mg/l, which is within the range associated with fatal paraquat poisoning.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,