Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1216818 | Journal of Chromatography B | 2006 | 6 Pages |
Abstract
A sensitive assay for the determination of unbound ertapenem in human plasma and bronchoalveolar lavage (BAL) was developed using ultrafiltration of plasma and BAL samples. A rapid HPLC method was used with ultraviolet detection set at a wavelength of 305 nm and a separation on a Prontosil AQ C18 column, with imipenem used as internal standard. This assay was linear over the concentration range of 0.5–100 μg/mL and 0.25–50 μg/mL in plasma and BAL, respectively. Limits of detection and quantitation were respectively 0.05 and 0.25 μg/mL. Validation data for accuracy and precision were CV < 2.48 and 8.25%, accuracy in the range 98.1–104.2% and 102.2–108.4%, respectively, for intra and inter-day.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Jean-Baptiste Gordien, Emmanuel Boselli, Catherine Fleureau, Bernard Allaouchiche, Gérard Janvier, Olivier Lalaude, Marie-Claude Saux, Dominique Breilh,