Article ID Journal Published Year Pages File Type
1216864 Journal of Chromatography B 2008 6 Pages PDF
Abstract

A new method for the determination of nine toxaphene specific congeners in fish liver oil and feedingstuff has been developed. The samples were extracted using pressurized liquid extraction followed by a purification on silica and florisil columns. Identification and quantification were conducted using GC–(EI)-HRMS, and comparison with MS/MS detection was performed, using electron ionization and negative chemical ionization. Limits of detection were ranged from 0.01 to 0.22 μg kg−1 (12% moisture) as required for feed samples. The calibration curves showed a good linearity for all congeners (R2 > 0.99). Repeatability was below 9% for all the congeners and recoveries were in-between 73 and 86%. This analytical method was applied to the quantification of thirteen real samples collected within national monitoring plans for further risk assessment.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,