Article ID Journal Published Year Pages File Type
1216913 Journal of Chromatography B 2012 5 Pages PDF
Abstract

A rapid and sensitive ultra-high performance liquid chromatography tandem mass spectrometric method was developed for simultaneous quantification of amoxicillin and prednisolone in bovine milk. In this method, amoxicillin, prednisolone and the internal standards penicillin G-d7 (for amoxicillin) and prednisolone-d6 were extracted from bovine milk using acetonitrile. The C18 solid phase extraction cartridges were selected for cleaning-up the extracts. The analytes were determined using a triple quadrupole mass spectrometry in positive electrospray ionization and multiple reaction monitoring mode. Calibration curves were linear over a concentration range of 2–1000 μg/kg for the analytes. The mean recoveries were 89.2–92.3% for amoxicillin and 98.7–102.3% for prednisolone. Limits of detection were 0.5 μg/kg for the analytes, and the limits of quantitation were 2 μg/kg. Decision limit (CCα) and detection capability (CCβ) have also been estimated for each analyte. The method was validated according to the Commission Decision 2002/657/EC and successfully applied to the analysis of amoxicillin and prednisolone in real samples.

► Amoxicillin and prednisolone were determined simultaneously using UPLC–MS/MS. ► Deuterated internal standards were applied to accurate quantification. ► LODs for two analytes were far below MRLs set by EU in developed method. ► Recoveries in raw milk were good with a simple sample preparation.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry