Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1217038 | Journal of Chromatography B | 2012 | 5 Pages |
Measurement of endogenous epinephrine (E) and norepinephrine (NE) in human plasma is very challenging due to lower endogenous concentrations as compared with animal plasma. An LC–MS/MS in combination with alumina-based SPE and derivatization procedure was validated for the measurement of E and NE in human plasma with acceptable intra-day and inter-day accuracy and precision. Sample was extracted with semi-automated alumina 96-well solid phase extraction (SPE) cartridge. The resulting eluent was dried and derivatized using d4-acetaldehyde. The analytes were separated on a monolithic C18 column. Extraction efficiencies were >66% for E and NE. The lower limit of quantitation (LLOQ) was 5.00 pg/mL for E and 20.0 pg/mL for NE.
► Measuring epinephrine and norepinephrine in human plasma. ► High throughput with semi-automated alumina-based SPE and derivatization procedure. ► Ultra sensitivity LC–MS/MS clinical biomarker assay.