Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1217173 | Journal of Chromatography B | 2007 | 6 Pages |
Abstract
A liquid chromatography-electrospray ionization-mass spectrometry (LC-ESI-MS) assay for the determination of bencycloquidium bromide (BCQB) in rat plasma was firstly developed and validated. After addition of 1-ethyl-bencycloquidium bromide as an internal standard (I.S.), the plasma samples were deproteinized with methanol and the supernatant was assayed by LC-ESI-MS. Chromatographic separation was achieved with a Hanbon Lichrospher 5-C18 column. The mobile phase consisted of methanol-40 mM ammonium acetate buffer-formic acid (75:25:0.25, v/v/v) and delivered at the flow rate of 1.0 ml/min. LC-ESI-MS was carried out on a single quadrupole mass spectrometer using electrospray ionization (ESI) and positive selected-ion monitoring (SIM). Target ions were monitored at [M]+m/z 330.2 for BCQB and [M] +m/z 344.2 for I.S. Calibration curve was linear over the range of 3-1500 ng/ml. The lower limit of quantification (LLOQ) was 3.0 ng/ml. The intra- and inter-run relative standard deviations (R.S.D.%) of the assay were less than 7.1 and 12.3%, respectively. The accuracy determined at the concentrations of 3.0, 100.0, 500.0 and 1500 ng/ml for BCQB were within ±15.0%. The established method has been applied successfully to study the pharmacokinetics of BCQB in rats after intranasal administration.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Qin Xu, Li Ding, Wenying Liu, Xiaojie Bian, Wenming Tang,