Article ID Journal Published Year Pages File Type
1217462 Journal of Chromatography B 2006 7 Pages PDF
Abstract

Succinylacetone (SA) is a specific marker for the inherited metabolic disease, hepatorenal tyrosinemia. We developed a stable-isotope dilution liquid chromatography tandem mass spectrometry for the determination of SA in dried blood spots (DBS) and liquid urine using a 13C4-SA as internal standard. SA was extracted, converted to the butyl ester and derivatized with dansylhydrazine (Dns-H). Calibration curves in DBS and urine calibrators were linear up to 100 and 30 μM, respectively. At a signal-to-noise ratio of 3, the limits of detection in DBS and urine were 0.2 and 0.005 μM, respectively. Total run time was 5 min. Intra- and inter-assay precision expressed as coefficient of variation were better than 9.1% with more than 96% recovery. The method was applied retrospectively and prospectively for the diagnosis of hepatorenal tyrosinemia and for follow-up of patients under treatment.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , , , ,