Article ID Journal Published Year Pages File Type
1217512 Journal of Chromatography B 2010 7 Pages PDF
Abstract

Remoxipride is a selective dopamine D2 receptor antagonist, and useful as a model compound in mechanism-based pharmacological investigations. To that end, studies in small animals with serial sampling over time are needed. For these small volume samples currently no suitable analytical methods are available. We propose analytical methods for the detection of low concentrations remoxipride in small sample volumes of plasma, brain homogenate, and brain microdialysate, using online solid phase extraction with liquid chromatography–tandem mass spectrometry. Method development, optimization and validation are described in terms of calibration curves, extraction yield, lower limit of quantification (LLOQ), precision, accuracy, inter-day- and intra-day variability. The 20 μl plasma samples showed an extraction yield of 76%, with a LLOQ of 0.5 ng/ml. For 0.6 ml brain homogenate samples the extraction yield was 45%, with a LLOQ of 1.8 ng/ml. The 20 μl brain microdialysate samples, without pre-treatment, had a LLOQ of 0.25 ng/ml. The precision and accuracy were well within the acceptable 15% range. Considering the small sample volumes, the high sensitivity and good reproducibility, the analytical methods are suitable for analyzing small sample volumes with low remoxipride concentrations.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , ,