Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1217580 | Journal of Chromatography B | 2008 | 5 Pages |
Abstract
A rapid, sensitive and specific method was developed and validated using LC/MS/MS for determination of sunitinib in human plasma. Sample preparation involved a liquid-liquid extraction by the addition of 0.2 mL of plasma with 4.0 mL tert-butyl-methyl-ether extraction solution containing 25 ng/mL of the internal standard clozapine. Separation of compounds was achieved on a C18 (50 mm Ã 2.1 mm i.d., 3.5 μm) analytical column using a mobile phase consisting of acetonitrile/H20 (65:35, v/v) containing 0.1% formic acid and isocratic flow at 0.150 mL/min for 3 min. The analytes were monitored by tandem-mass spectrometry with electrospray positive ionization. Linear calibration curves in human plasma were generated over the range of 0.2-500 ng/mL with values for the coefficient of determination of >0.9950. Within- and between day precision and accuracy were â¤10%. The method was applied to the quantitation of sunitinib in plasma samples from a patient receiving daily oral therapy with sunitinib.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Patton Minkin, Ming Zhao, Zhaoyuan Chen, Jan Ouwerkerk, Hans Gelderblom, Sharyn D. Baker,