Abstract
A simple RP-HPLC method based on fluorescence detection was developed for the quantitation of 7-amino-4-trifluoro methylcoumarin (AFC) in cell lysates from JEG-3 choriocarcinoma cells for determination of caspase-4 activity. In contrast to the established methods of AFC detection using a fluorescence microplate reader or using a fluorescence photometer, the separation of AFC-signals from interfering fluorescence signals by a reversed phase column affords more precise quantitation of released AFC. This can be important for analyses of cell lysates with low caspase activity or experimental series with marginal differences among samples. By applying this new method, a linear dynamic range of 40 pmol/mL to 3 nmol/mL with a correlation coefficient of 0.9996 was achieved. Due to the short retention time (â¼7 min), the determination of AFC by RP-HPLC under isocratic conditions requires small amounts of samples (50 μL injection volume), and allows increased sample throughput. This method should be easily applied with little or no modification to other caspase assays by using the same fluorophore.
Keywords
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Sandra Koehn, Mike Trueck, Tobias G. Poehlmann, Ekkehard Schleussner, Udo R. Markert, Lydia Seyfarth,