Abstract

•The parameters of filter-furnace for atomic absorption spectroscopy are presented.•Using Pd-Mg chemical modifier.•The background absorption signal was successfully reduced.•The time of one element determination does not exceed ∼3–5 min.•Determination trace amounts of heavy metals in some of vegetable oils and fats.

The article presents the results of optimization of operation parameters, investigation of analytical characteristics and the abilities of a graphite filter-furnace (FF) atomizer for the direct electrothermal atomic absorption spectroscopy (ET AAS) determination of trace amounts of Mn, As, Pb, Cu, Cd and Zn in some vegetable oils and fats. The effect of pyrolysis and atomization temperatures of the graphite FF atomizer on atomic absorbance values of the listed elements at their evaporation from some organic solutions in the presence of a Pd-Mg chemical modifier (CM) was investigated. For the ET AAS determination of As, Pb, Cd and Zn with Pd-Mg CM, the temperature of the graphite FF atomizer for the pyrolysis step can be raised by 250–350 °C. This mode allows to eliminate the background absorption, to increase the sensitivity of the elements to be analyzed and to enhance the total content of vegetable oils or fats in organic solutions up to 0.5 g mL−1. The obtained limits of quantification for Mn, As, Pb, Cu, Cd and Zn were 0.002, 0.004, 0.004, 0.002, 0.0008, 0.0004 mg kg−1, respectively. The relative standard deviation (RSD) varied between ∼3 and 8% and the time of one element determination did not exceed ∼3–5 min. The reliability of the proposed method was checked using the reference method. A paired Student's t-test showed no significant difference between the results obtained by both methods on a 95% confidence level.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,