Article ID Journal Published Year Pages File Type
1220819 Journal of Pharmaceutical and Biomedical Analysis 2015 8 Pages PDF
Abstract

•Simultaneous determination of sofosbuvir and GS-331007 in plasma was done.•The method was applied for a bioequivalence study.•A validated, highly sensitive UPLC-ESI-MS/MS method was developed.

A rapid and simple LC-MS/MS method was developed and validated for the simultaneous estimation of sofosbuvir (SF) and its metabolite GS-331007 (GS) using famotidine as an internal standard (IS). The Xevo TQD LC-MS/MS was operated under the multiple-reaction monitoring mode using electrospray ionization. Extraction with ethyl acetate was used in sample preparation. The prepared samples were chromatographed on Acquity UPLC HSS C18 (50 mm × 2.1 mm, 1.8 μm) column by pumping 0.1% formic acid and acetonitrile (50:50, v/v) in an isocratic mode at a flow rate of 0.3 ml/min. Method validation was performed as per the FDA guidelines and the standard curves were found to be linear in the range of 10–2500 ng/ml for both SF and its metabolite. The intra-day and inter-day precision and accuracy results were within the acceptable limits. A very short run time of 1.2 min made it possible to analyze more than 300 human plasma samples per day. The developed assay method was successfully applied to a bioequivalence study in human volunteers.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,