Article ID Journal Published Year Pages File Type
1220939 Journal of Pharmaceutical and Biomedical Analysis 2016 9 Pages PDF
Abstract

•18 bioactive compounds of six species were analyzed under different DART operating parameters.•DART-MS analytical results were evaluated by peak area and S/N of the characteristic ion signals.•The relationships among compounds, DART conditions and analytical responses were clarified.•Adequate ion source operating parameters were proposed for types of TCM active substance.

Direct analysis in real time mass spectrometry (DART-MS) provides a new analytical method for traditional Chinese medicine (TCM). The present study investigated the effects of key ion source operating parameters on DART-MS analysis of various TCM active components. A total of 18 active components, including phenylpropanoids, alkaloids, saponins, flavones, volatile oils, and glycosides, were examined. For each substance, the peak area and signal-to-noise of its characteristic ions under different reagent gases and heater temperatures were compared. Based on the comparison, the relationships among chemical structures, ion source parameters and instrument responses were revealed. Finally, some suggestions about choosing reagent gas and heater temperature were proposed for types of TCM active substance, which offered a reference for the application of DART-MS on TCM analysis.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,