Abstract

The objective of this work was to develop and validate a rapid reversed-phase (RP) high-performance liquid chromatography (HPLC) method for the in vitro pharmaceutical characterization of dapivirine-loaded polymeric nanoparticles. Chromatographic runs were performed on a RP C18 column with a mobile phase comprising acetonitrile–0.5% (w/v) triethanolamine solution in isocratic mode (80:20, v/v) at a flow rate of 1 ml/min. Dapivirine was detected at a wavelength of 290 nm. The method was shown to be specific, linear in the range of 1–50 μg/ml (R2 = 0.9998), precise at the intra-day and inter-day levels as reflected by the relative standard deviation values (less than 0.85%), accurate (recovery rate of 100.17 ± 0.35%), and robust to changes in the mobile phase and column brand. The detection and quantitation limits were 0.08 and 0.24 μg/ml, respectively. The method was successfully used to determine the loading capacity and association efficiency of dapivirine in poly(lactic-co-glycolic acid)-based nanoparticles and its in vitro release.