Article ID Journal Published Year Pages File Type
1222186 Journal of Pharmaceutical and Biomedical Analysis 2012 6 Pages PDF
Abstract

The quantitative analysis of glucose, triglycerides and high-density lipoprotein (HDL) in rat plasma without sample pre-treatment using direct near-infrared spectroscopy was studied. Comparison was made of several multivariate calibration techniques and algorithms for data pre-processing and variable selection, including partial least squares (PLS), interval partial least squares (iPLS), genetic algorithm (GA) and successive projections algorithm (SPA). Variable selection yielded good results for the correlation coefficient and Root Mean Square Error of Prediction (RMSEP) values for the three parameters, especially triglycerides. The RMSEP values for glucose, triglycerides and HDL produced by the PLS model were 6.08, 16.07 and 2.03 mg dl−1, respectively. F tests and t-tests were performed to compare the results of the models with each other and with a reference method. These results suggests that the PLS method can be used to simultaneously determine the concentrations of glucose, triglycerides and HDL in complicated biological fluids with NIR spectroscopy, offering an alternative analysis in animals.

► NIRS was applied the simultaneous determination glucose, triglycerides and high-density lipoprotein (HDL) in rat plasma. ► To improve determination models variable selection methods (iPLS, GA, SPA) were evaluated. ► The RMSEP values for glucose, triglycerides and HDL produced by the PLS model were 6.08, 16.07 and 2.03 mg dl−1, respectively.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,