Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1222187 | Journal of Pharmaceutical and Biomedical Analysis | 2012 | 6 Pages |
A high performance liquid chromatography–electrospray ionization-tandem mass spectrometric method (HPLC–ESI-MS/MS) was developed for simultaneous determination of adenosine 5′-triphosphate (ATP), adenosine 5′-diphosphate (ADP), adenosine 5′-monophosphate (AMP), creatine phosphate (CP), and creatine in rat liver. After extraction with pre-cooled (4 °C) methanol/water (1:1, v/v), the analytes were separated on a porous graphitic carbon (Hypercarb) column (2.1 mm × 150 mm, 5 μm) using a programmed gradient elution with a mobile phase consisting of 2 mmol/L ammonium acetate in water and 2 mmol/L ammonium acetate in acetonitrile (pH = 10.0). The analytes were detected in a way of multiple reaction monitoring (MRM) under negative scan mode by a triple quadrupole mass spectrometer with electrospray ionization (ESI). An external calibration method with linear ranges from 10 to 5000 ng/mL for the five target compounds was used for quantification with a correlation coefficients ≥ 0.9973. The limits of detection and limits of quantification for all analytes were in ranges from 0.50 to 1.5 ng/mL and 1.6 to 0.5 ng/mL, respectively. The average recoveries spiked in three levels were from 77.2% to 102% and precisions expressed in RSDs were from 0.2% to 4.8%. The established method was successfully applied to determination of ATP, ADP, AMP, CP and creatine in liver tissue.
► A simple and reliable HPLC–ESI-MS/MS method was developed for simultaneous determination of ATP, ADP, AMP, CP and creatine in liver tissues. ► The inorganic buffer and/or ion-pairing reagents were avoided and reasonable retention time for the analytes was achieved. ► The sample pre-treatment procedure was rapid, simple and well compatible with ESI-MS/MS analysis. ► The method was successfully applied to determination of ATP, ADP, AMP, CP and creatine in liver tissue.