Article ID Journal Published Year Pages File Type
1222312 Journal of Pharmaceutical and Biomedical Analysis 2012 5 Pages PDF
Abstract
The present work describes reversed-phase high performance liquid chromatographic methodologies (RP-HPLC) for the qualitative and quantitative analysis of the human glycoprotein hormones thyrotropin (hTSH), follitropin (hFSH), choriogonadotropin (hCG) and lutropin (hLH) in the presence of a large excess (up to 250:1) of human serum albumin (HSA). Chromatographic profiles with a good separation between the hormone and HSA were obtained by using a C4 column and specific gradient elution conditions for each hormone. Parameters such as resolution factor, tailing factor and relative retention time, were determined, and are useful for the evaluation of the quality of the separation obtained between the active pharmaceutical ingredient and the excipient present in the final formulation. The potential of each method for quantification of both HSA and the hormone was also demonstrated. Besides furnishing chromatographic quantifications that can substitute for in vivo bioassays and animal use, the chromatograms also provide a direct panorama of the quality and heterogeneity of the protein of interest.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,