Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1223134 | Journal of Pharmaceutical and Biomedical Analysis | 2011 | 8 Pages |
Abstract
The aim of this work was to develop and validate more suitable sample preparation method for the determination of atorvastatin and its metabolites in biological samples using MEPS (microextraction by packed sorbent). The optimal conditions of MEPS extraction were using C8 sorbent and only 50 μl of the sample. The analytes were eluted by 100 μl of the mixture of acetonitrile:0.1 M ammonium acetate pH 4.5 (95:5, v:v). The analytical method was validated and demonstrated good linearity (r2 > 0.9990), recovery (89-115%) and intra-day precision (RSD < 10%). Total time of the sample preparation was three times shorter (7 min) compared to SPE. The volume of sample was twenty times lower and the volume of solvents about ten times lower compared to SPE. Combination of fast MEPS method together with quick UHPLC-MS/MS was used for the determination of atorvastatin and its two metabolites in serum obtained from patients with familiar hypercholesterolemia.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Hana VlÄková, Dagmar Solichová, Milan Bláha, Petr Solich, Lucie Nováková,