Article ID Journal Published Year Pages File Type
1223211 Journal of Pharmaceutical and Biomedical Analysis 2010 7 Pages PDF
Abstract

A rapid and sensitive method was developed and validated using a liquid chromatographic method with tandem mass spectrometry detection (LC/MS/MS) for determination of veliparib (ABT-888) in plasma, bone marrow supernatant, and bone marrow cells. Sample preparation involved a single protein precipitation step by the addition of the sample with acetonitrile. Separation of veliparib and the internal standard, A620223.69, was achieved on a Atlantis™ dC18 column (100 mm × 2.1 mm, 3 μm) column using a mobile phase consisting of acetonitrile–ammonium acetate (2 mM) containing formic acid (0.1%, v/v) using isocratic flow at 0.2 mL/min for 3 min. The analyte and internal standard were monitored by tandem mass spectrometry with electrospray positive ionization. Linear calibration curves were generated over the range of 5–1000 nM. The values for both within day and between day precision and accuracy were well within the generally accepted criteria for analytical methods. This method was subsequently used to measure concentrations of veliparib in cancer patients receiving an oral daily dose of 10 mg with demonstration of drug accumulation in the marrow compartment and in the target leukemia bone marrow cells.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , ,