Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1223808 | Journal of Pharmaceutical and Biomedical Analysis | 2006 | 6 Pages |
Packed-column supercritical fluid chromatography (pSFC) coupled to an atmospheric pressure chemical ionization (APCI) source and a tandem mass spectrometer (MS/MS) for rapid and simultaneous determination of clozapine, ondansetron, tolbutamide and primidone in in vitro samples was developed in support of metabolic stability experiments. The effects of the eluent flow-rate and composition as well as the nebulizer temperatures on the ionization efficiency of the analytes in positive ion mode under normal phase pSFC conditions were studied. The metabolic stability of the test drug components through microsomal incubation by the proposed pSFC–APCI/MS/MS approaches requiring approximately 1 min per samples were evaluated with respect to specificity, durability and accuracy. These metabolic stability results obtained by pSFC–MS/MS methods are in a good agreement with those obtained by fast high-performance liquid chromatography/tandem mass spectrometry (HPLC–MS/MS) method.