Article ID Journal Published Year Pages File Type
1224456 Journal of Pharmaceutical and Biomedical Analysis 2007 7 Pages PDF
Abstract

The design and validation of a novel high-throughput system for thermodynamic solubility determination requiring only 5 mg of sample is described. The system uses a sintered nickel filter assembly to recover excess solids from saturated solutions for rapid crystallinity assessment via powder X-ray diffraction (PXRD). Moreover, the system measures the pH of filtrates to provide a final pH value with the solubility measurement. The limit of detection for the UV–vis plate reader used on this system is ∼0.001 mg/ml, while the practical upper limit is ∼3 mg/mL. The solubility measurements of 60 proprietary Pfizer compounds were used to validate the nickel filter assembly against a more conventional polyvinylidenedifluoride (PVDF) filter. Additionally, a comparison was made between a subset of 10 compounds run on the automated system and a more traditional shake-flask method employing HPLC analysis. In both cases, a favorable comparison was obtained.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,