Article ID Journal Published Year Pages File Type
1226511 Journal of Trace Elements in Medicine and Biology 2014 6 Pages PDF
Abstract

Cloud point extraction (CPE) was used to simultaneously preconcentrate trace-level cadmium, nickel and zinc for determination by flame atomic absorption spectrometry (FAAS). 1-(2-Pyridilazo)-2-naphthol (PAN) was used as a complexing agent, and the metal complexes were extracted from the aqueous phase by the surfactant Triton X-114 ((1,1,3,3-tetramethylbutyl)phenyl-polyethylene glycol). Under optimized complexation and extraction conditions, the limits of detection were 0.37 μg L−1 (Cd), 2.6 μg L−1 (Ni) and 2.3 μg L−1 (Zn). This extraction was quantitative with a preconcentration factor of 30 and enrichment factor estimated to be 42, 40 and 43, respectively. The method was applied to different complex samples, and the accuracy was evaluated by analyzing a water standard reference material (NIST SRM 1643e), yielding results in agreement with the certified values.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,