Article ID Journal Published Year Pages File Type
1227555 Microchemical Journal 2016 6 Pages PDF
Abstract

•A FI-SPE-UV system has been designed for Pb determination.•Oxidized carbon xerogel is introduced for the first time in online SPE procedures.•Oxidized carbon xerogel showed quantitative Pb retention.•Good Pb kinetic desorption from carbon xerogel allowed high enrichment factors.•Method sensibility was compatible with maximum Pb concentrations in waters established by WHO.

The proposed methodology is based on the preconcentration of Pb2 + on oxidized carbon xerogel, a porous material of low density and high surface area. Adsorbed Pb2+ was eluted with HCl, followed by complexion with dithizone in ethanol–water solution and determined by UV–visible spectrophotometry in a continuous flow system (FI). Variables such as pH, type of eluent, eluent concentration, type of complexing solvent, concentration of complexing reagent and flow rates were optimized. The proposed system reached an enrichment factor of 50, with 5 mL of sample. Detection limit was of 0.9 μg L− 1, with a precision of 7.8% (n = 10). The proposed system was successfully applied to the determination of Pb in tap water from San Juan City, Argentina.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,