Article ID Journal Published Year Pages File Type
1227842 Microchemical Journal 2010 4 Pages PDF
Abstract

In this work, the determination of total As in seawater by hydride generation atomic fluorescence spectrometry was studied. The influence of the chemical, flow and instrumental parameters were investigated and optimized. The pre-reduction of As(V) to As(III) was performed using KI plus ascorbic acid in 3.5 mol L− 1 HCl medium. No multiplicative interference was present and external aqueous calibration could be used. The limit of detection was 36 ng L− 1, while the repeatability was 2% (n = 10), at a 500 ng L− 1 concentration level. The sample throughput was 15 h− 1 if triplicate measurements were made. The accuracy was assessed by the analysis of a seawater certified reference material and excellent agreement between the obtained and certified values was verified. The procedure was used for the analysis of seawater offshore samples collected at the Brazilian coast and results ranging from 860 to 1200 ng L− 1 were found.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry