Article ID Journal Published Year Pages File Type
1237083 Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 2009 6 Pages PDF
Abstract
Hydrophobic polymer layers (3 μm) were spin-coated on Si or Ge plates and placed in a flow through gas chamber. FTIR reflection spectra of the layers were recorded showing the characteristic IR absorption bands of the polymer and the interference pattern generated by layered structure of the polymer film. Upon exposure of the polymer layer to gaseous analytes enrichment in the polymer film occurred. This was evidenced by the appearance of analyte specific absorption particular in the mid-IR part of the spectrum, as well as by a shift in the interference pattern across the whole spectrum. Qualitative information concerning the analyte was accessible in the mid-IR part of the spectrum, whereas quantitative assessment was obtained from the interference pattern. Polyetherurethane, polydimethylsiloxane, Makrolon® and polyisobutylene polymer layers were tested for such IR-RIfS measurements, whereas toluene, o-dichlorobenzene, m-xylene, ethyl acetate and cyclohexane were employed as analytes. There was no influence of water vapour neither on the IR absorptions nor the interference pattern as hydrophobic polymers were used.
Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,