Article ID Journal Published Year Pages File Type
1238707 Spectrochimica Acta Part A: Molecular and Biomolecular Spectroscopy 2006 5 Pages PDF
Abstract

We have measured the output parameters of a 10.3-μm pulsed distributed-feedback (DFB) quantum cascade (QC) laser manufactured by Alpes Lasers and intended for high-sensitivity detection of ammonia and ethylene. The laser beam was collimated with an AR-coated aspheric ZnSe lens with focal length of 11.6 mm and clear aperture of 16.5 mm. Near- and far-field distributions of the laser emission were recorded with an infrared imaging camera. The fast-and slow-axis laser beam divergences were measured to be 1.2 and 1.4 mrad (FWHM), respectively. The divergence was found to be increasing with injection current. An air-spaced Fabry–Perot interferometer with free spectral range of 0.05 cm−1 was used to measure the frequency tuning rates of the laser. The laser was tuned by either heat sink temperature, injection current or pulse repetition rate with rates of ∼−8 × 10−2 cm−1 K−1, −7 × 10−2 cm−1 A−1 and −9 × 10−4 cm−1 kHz−1, respectively. The laser frequency decreased linearly with a rate of 10−2 cm−1 ns−1 (∼300 MHz ns−1) for laser pulses varied from 10 to 50 ns, and the frequency chirp rate was found to decrease for longer laser pulses.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , ,