Article ID Journal Published Year Pages File Type
1239885 Spectrochimica Acta Part B: Atomic Spectroscopy 2012 6 Pages PDF
Abstract

The frequency response of a free running RF generator was evaluated for its ability to provide plasma diagnostic information. A small probe coil in the plasma torch housing and a frequency counter were used for frequency measurement. A linear dependence between the RF generator frequency and the flow of a foreign gas (CO2, H2, He, N2—introduced into the central channel of the ICP) was encountered. Moreover, the inductively coupled plasma (ICP) was found to extinguish at a constant, characteristic frequency when the flow of foreign gas was too high, allowing a simple prediction of the maximal tolerable foreign gas flow. Spectroscopic plasma diagnostic techniques (Mg II/Mg I ratio, excitation temperature and rotational temperature) were also investigated but provided little indication of instable plasma conditions due to the addition of foreign gases.

► Frequency response of a free running RF generator for plasma diagnostics. ► Linear dependence of the RF generator frequency on foreign gas flow (CO2, H2, He, N2). ► ICP extinction frequency is independent from the foreign gas.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,