Abstract

A new wavelength-dispersive X-ray fluorescence (WD-XRF) imaging spectrometer equipped with a high-sensitivity imaging sensor was developed in our laboratory. In this instrument, a straight polycapillary optic was applied instead of a Soller slit as well as a 2D imaging X-ray detector instead of X-ray counters, which are used in conventional WD-XRF spectrometers. Therefore, images of elemental distribution were available after a short exposure time. Ni Kα images and Cu Kα images were clearly obtained at corresponding diffraction angles for a short exposure time of 10 s. By optimizing the spectrometer, the time required for imaging is reduced, leading to XRF image movies. It is difficult to distinguish two peaks (Ti Kα (4.508 keV) and Ba Lα (4.465 keV)) due to the poor energy resolution of EDXRS. However, Ti and Ba images could be successfully observed by the WD-XRF imaging spectrometer. The energy resolution of the developed spectrometer was 25 eV at the Ti Kα peak.

► We developed a wavelength dispersive X-ray fluorescence imaging spectrometer. ► A high-sensitivity sensor (PILATUS) was applied for X-ray elemental imaging. ► WD-XRF images with a short exposure time of 10 s were demonstrated. ► A high energy-resolution (less than 40 eV) was achieved.