Abstract

Calibration is required to obtain analyte concentrations in atomic spectrometry. To take full benefit of it, the adequacy of the coefficient of determination r2 is discussed, and its use is compared with the uncertainty due to the prediction bands of the regression. Also discussed from a tutorial point of view are the influence of the weighting procedure and of different weighting factors, and the comparison between linear and quadratic regression to cope with curvatures. They are illustrated with examples based on the use of ICP-AES with nebulization and laser ablation, and of LIBS. Use of a calibration graph over several orders of magnitude may be problematic as well as the use of a quadratic regression to cope with possible curvatures. Instrument softwares that allow reprocessing of the calibration by selecting standards around the expected analyte concentration are convenient for optimizing the calibration procedure.