Article ID Journal Published Year Pages File Type
1240765 Spectrochimica Acta Part B: Atomic Spectroscopy 2007 6 Pages PDF
Abstract

In this work, the use of cetyltrimethylammonium bromide as surfactant for the preparation of oil-in-water emulsions for the determination of Cu and Cr in gasoline by electrothermal atomic absorption spectrometry (ET AAS) was evaluated. The surfactant amount was tested in the range of 25 to 300 mg, added to 2 ml of gasoline, and completed to 10 mL with 0.1% (v/v) nitric acid solution. 150 mg of surfactant was found optimum, and a sonication time of 10 min sufficient to form an oil-in-water emulsion that was stable for several hours. The ET AAS temperature program was established based on pyrolysis and atomization curves. The pyrolysis temperatures were set at 700 and 1300 °C for Cu and Cr, respectively and the selected atomization temperatures were 2400 and 2500 °C. The time and temperature of the drying stage and the atomization time were experimentally tested to provide optimum conditions. The limits of detection were found to be 5 μg L− 1 and 1.5 μg L− 1 for Cu and Cr, respectively in the original gasoline samples. The relative standard deviation (RSD) ranged from 4 to 9% in oil-in-water emulsions spiked with 5 μg L− 1 and 15 μg L− 1 of each metal, respectively. Recoveries varied from 90 to 98%. The accuracy of the proposed method was tested by an alternate procedure using complete evaporation of the gasoline sample. The method was adequate for the determination of Cu and Cr in gasoline samples collected from different gas stations in Salvador, BA, Brazil.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,