Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1240841 | Spectrochimica Acta Part B: Atomic Spectroscopy | 2006 | 5 Pages |
Abstract
Within the framework of a collaborative project, it is shown that commercial total reflection X-ray fluorescence (TXRF) systems used in laboratories can easily be upgraded with a silicon drift detector (SDD). SDDs have advantages when used with fully automatized wafer analyzers working under cleanroom conditions, because no liquid nitrogen is required as they are electrically cooled. The goal of this work was the integration of a KETEK 10Â mm2 SDD in an ATOMIKA 8030W wafer analyzer with special attention to maintain the high degree of automation of the system. An electronic device was designed to establish communication between the SDD and the TXRF electronic control system. The adapted system was tested and compared with the original setup using an 80Â mm2 Si(Li) detector. Multielement droplet samples on silicon wafers were analyzed and the results showed two times better detection limits for the Si(Li) detector for 1000Â pg Ni in comparison to the SDD. Additionally, a RADIANT 50Â mm2 SDD (VORTEX) was tested which showed identical detection limits compared to the 80Â mm2 Si(Li) detector.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Siegfried Pahlke, Florian Meirer, Peter Wobrauschek, Christina Streli, Georg Peter Westphal, Claus Mantler,