Article ID Journal Published Year Pages File Type
1241460 Spectrochimica Acta Part B: Atomic Spectroscopy 2006 4 Pages PDF
Abstract

Conventional and microwave assisted digestion, both using aqua regia, alkaline fusion with lithium metaborate and aqueous slurries were evaluated as sample treatments for determination of Pt in automotive catalytic converters by Graphite Furnace Atomic Absorption Spectrometry (GF-AAS). Determination of platinum by GF-AAS in samples of the catalytic converter's substrates, prepared by the four methods described, indicates that the highest platinum concentration i.e. maximum Pt extraction in the range of 748 ± 15–998 ± 10 μg mL− 1, is obtained for samples dissolved by alkaline fusion, closely followed by analysis of aqueous plus Triton X-100 slurries 708 ± 14–958 ± 10 μg mL− 1, while neither one of the acid digestion procedures achieved total dissolution of the samples. Slurry analysis is thus shown to be a viable alternative and is recommended, based on its speed and ease of implementation. Aqueous standards calibration curves and the standard addition methods were also compared. The results showed that no appreciable matrix effects are present, regardless of the sample preparation procedure used. Precision of the measurements, expressed as percentage relative standard deviation, ranged between 2.5 to 4.9%. Accuracy of the results was assessed by recovery tests which rendered values between 98.9 and 100.9%.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,