Article ID Journal Published Year Pages File Type
1242364 Talanta 2011 5 Pages PDF
Abstract

Poly(vinyl chloride) (PVC)-based membranes of N,N-bis-succinamide-based dendritic molecule with tetrabutyl ammonium bromide (TBAB) as a cation inhibitor and dibutylphthalate (DBP), dioctylphthalate (DOP), dibutyl (butyl) phosphonate (DBBP) and 1-chloronaphthalene (CN) as plasticizing solvent mediators were prepared and used as nitrate ion-selective electrodes. Optimum performance was observed with the membrane having I–PVC–TBAB–DBP in the ratio 1:33:1:65 (w/w). The electrode has a linear response to nitrate with a detection limit of 3.9 × 10−5 ± 0.07 M and Nernstian compliance (57.0 ± 0.2 mV/decade) between pH 2.8 and 9.6 with a fast response time of about 20 s. The selectivity coefficient values of the order of 0.001 for mono-; bi- and trivalent anions; indicate high selectivity for nitrate ions over these anions. The preparation procedure of the electrode is very easy and inexpensive. The electrodes were used over a period of 45 days with good reproducibility. The analytical usefulness of the proposed electrode has been evaluated by its application in the determination of nitrate ions in waste water samples.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , ,