Abstract

•Polymer macroporous platforms can be used for detection of nucleotide replacements.•The diagnostic test system on several genomic substitutions was developed.•High sensitivity and specificity of developed test system was demonstrated.•High accuracy of developed biochip in detection of genomic substitutions was shown.

Analysis of variations in DNA structure using a low-density microarray technology for routine diagnostic in evidence-based medicine is still relevant. In this work the applicability of 3-D macroporous monolithic methacrylate-based platforms for detection of different pathogenic genomic substitutions was studied. The detection of nucleotide replacements in F5 (Leiden G/A, rs6025), MTHFR (C/T, rs1801133) and ITGB3 (T/C, rs5918), involved in coagulation, and COMT (C/G, rs4818), TPH2 (T/A, rs11178997), PON1 (T/A rs854560), AGTR2 (C/A, rs11091046) and SERPINE1 (5G/4G, rs1799889), associated with pregnancy complications, was performed. The effect of such parameters as amount and type of oligonucleotide probe, amount of PCR product on signal-to-noise ratio, as well as mismatch discrimination was analyzed. Sensitivity and specificity of mutation detections were coincided and equal to 98.6%. The analysis of SERPINE1 and MTHFR genotypes by both NGS and developed microarray was performed and compared.

Graphical abstractFigure optionsDownload full-size imageDownload as PowerPoint slide