Article ID Journal Published Year Pages File Type
1243128 Talanta 2008 4 Pages PDF
Abstract

A new micelle-mediated cloud point extraction method is described for sensitive and selective determination of trace amounts of rhodamine B by spectrophotometry. The method is based on the cloud point extraction of rhodamine B from aqueous solution using Triton X-100 in acidic media. The extracted surfactant rich phase is diluted with water and its absorbance is measured at 563 nm by a spectophotometer. The effects of different operating parameters such as concentration of surfactant and salt, temperature and pH on the cloud point extraction of rhodamine B were studied in details and a set of optimum conditions were obtained. Under optimum conditions a linear calibration graph in the range of 5–550 ng mL−1 of rhodamine B in the initial solution with r = 0.9991 (n = 15) was obtained. Detection limit based on three times the standard deviation of the blank (3Sb) was 1.3 ng mL−1 (n = 10) and the relative standard deviation (R.S.D.) for 50 and 350 ng mL−1 of rhodamine B was 2.40 and 0.87% (n = 10), respectively. The method was applied for the determination of rhodamine B in soft pastel, hand washing liquid soap, matches tip and textile dyes mixture samples.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,