Abstract

A simple and straightforward quantification method integrated in a fully automated multi-pumping flow system (MPFS) using water-soluble mercaptopropionic acid (MPA)-capped CdTe quantum dots (QDs) was implemented for the fluorescence quantification of N-acetyl-l-cysteine (NAC) in pharmaceutical formulations. The developed approach was based on NAC ability to establish surface interactions that result in enhanced nanocrystals fluorescence intensity, proportional to analyte concentration. Size and concentration of QDs, ageing, composition, concentration and pH of the buffer solution revealed to have a noticeable effect on the enhancing efficiency affecting sensitivity and linear working range of the methodology.Under the optimal conditions, a linear working range was obtained for NAC concentrations ranging from 50 to 750 μmol L−1 (r = 0.9978), with good precision (r.s.d. < 1.6%; n = 5) and a sampling rate of about 75 hr−1. The detection limit (LOD) was approximately 1.6 μmol L−1. The method was applied to pharmaceutical preparations and the results revealed good agreement with those obtained by the reference procedure with relative deviations between −2.1 and +4.2%. Advantages of the new procedure include speed, low consumption of reagents, minor waste generation, requiring also much less work than the recommended HPLC method. The mechanism for luminescence enhancement of CdTe QDs is discussed. FT-IR spectra revealed that sulphydryl groups of NAC have a high affinity with the nanocrystals.