Abstract

The highly sensitive determination of lead (Pb(II)) and cadmium (Cd(II)) ions, with a limit of detection of 0.01 μg L−1 for Pb(II) and Cd(II), by on-line preconcentration and anodic stripping voltammetry (ASV) controlled by a sequential injection analysis (SIA) system is reported here. The SIA system consisted of a syringe pump, an 8-port selection valve and a 6-port switching valve and was incorporated with a bismuth coated screen-printed carbon nanotube electrode (Bi-SPCNTE). The preconcentration of metal ions was performed by solid phase extraction using an Analig TE-05 chelating resin mini-column on a switching valve. The metal ions collected were then eluted from the resin with 1 M hydrochloric acid (HCl), deposited on the electrode surface at −1.3 V vs. Ag/AgCl and then measured with ASV. The pH of the sample, eluent volume, flow rate, concentration of the bismuth plating solution and the square-wave voltammetric parameters were optimized. Under the optimum conditions, an enrichment factor of 11.9-fold and 6.6-fold for Pb(II) and Cd(II) ions, respectively, was attained. Detection of Pb(II) and Cd(II) had two different linear ranges (0.5–15 μg L−1 and 15–70 μg L−1).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,