Article ID Journal Published Year Pages File Type
1243687 Talanta 2009 6 Pages PDF
Abstract

A sensitive and efficient method was developed for determination of metoprolol in human plasma by gas chromatography–mass spectrometry (GC–MS). Metoprolol and atenolol (internal standard, IS) were extracted from human plasma with a mixture of ethylacetate and diethylether at basic pH with liquid–liquid extraction. Calibration curves were linear over the concentration range 15–500 ng/ml. Intra- and inter-day precision values for metoprolol in human plasma were less than 6.4, and accuracy (relative error) was better than 8.8%. The analytical recovery of metoprolol from human plasma averaged 91.20%. The limits of detection (LOD) and quantification (LOQ) of metoprolol were 5.0 and 15 ng/ml, respectively. Also the developed and validated GC–MS method was successfully applied to three patients with hypertension who had been given an oral tablet of 100 mg metoprolol.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,