Article ID Journal Published Year Pages File Type
1243888 Talanta 2008 6 Pages PDF
Abstract

In this work, an amperometric sensor for nitrite detection based on a glassy carbon electrode modified with copper tetrasulphonated phthalocyanine immobilized by polycationic poly-l-lysine film is presented. The modified electrode showed an excellent catalytic activity toward nitrite oxidation. A linear response range from 0.12 up to 12.20 μmol L−1 was obtained with a sensitivity of 0.83 μA L μmol−1. The detection limit for nitrite was 36 nmol L−1. The repeatability of the proposed sensor, evaluated in terms of relative standard deviation, was 1% for 10 measurements of 10 μmol L−1 nitrite solution. Finally, the developed sensor was applied for nitrite determination in water samples and the results were in agreement to the comparative method. The average recovery for the samples was 101 (±4)%.