Article ID Journal Published Year Pages File Type
1243963 Talanta 2008 8 Pages PDF
Abstract

For the first time, a multisyringe flow injection analysis capillary electrophoresis system is described. The potential of the hyphenation for sample treatment including analyte pre-concentration is demonstrated by its successful application to the determination of mono-nitrophenols (NPs) in different water samples.The analytical system was used to automate in-line sample acidification, analyte pre-concentration, elution, hydrodynamic injection, electrophoretic separation, and detection as well as the maintenance and re-conditioning of the solid-phase extraction (SPE) column and the separation capillary.A pre-concentration factor of better than 115 and detection down to 0.11 μmol L−1 were achieved. Detection was carried out at visible wavelength using a blue LED as a low baseline-noise light source. High repeatability was obtained each for migration times and for peak heights with relative standard deviations typically below 2.5 and 6% including the pre-concentration procedure, respectively.Three injections per hour were achieved by running in parallel the pre-concentrating and the electrophoretic separation procedures. Instrumental control and data registration and evaluation were carried out with the software package AutoAnalysis, allowing autonomous operation of the analytical system.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , ,