Article ID Journal Published Year Pages File Type
1244172 Talanta 2011 13 Pages PDF
Abstract

A new, simple, economical and validated high-performance liquid chromatography linked with electrochemical detector (HPLC–ECD) method has been developed and optimized for different experimental parameters to analyze the most common monothiols and disulfide (cystine, cysteine, homocysteine, methionine, reduced (GSH) and oxidized glutathione (GSSG)) and ascorbic acid present in human plasma and erythrocytes using dopamine as internal standard (IS). Complete separation of all the targets analytes and IS at 35 °C on Discovery HS C18 RP column (250 mm × 4.6 mm, 5 μm) was achieved using 0.05% TFA:methanol (97:3, v/v) as a mobile phase pumped at the rate of 0.6 ml min−1 using electrochemical detector in DC mode at the detector potential of 900 mV. The limits of detection (3 S/N) and limits of quantification (10 S/N) of the studied compounds were evaluated using dilution method. The proposed method was validated according to standard guidelines and optimization of various experimental parameters and chromatographic conditions was carried out. The optimized and validated HPLC–ECD method was successfully applied for the determination of the abovementioned compounds in human plasma and erythrocytes. The method will be quite suitable for the determination of plasma and erythrocyte profile of ascorbic acid and aminothiols in oxidative stress and other basic research studies.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , , , , , , , ,