Article ID Journal Published Year Pages File Type
1244545 Talanta 2008 6 Pages PDF
Abstract

Chitosan membrane with glutathione reductase and sulfhydryl oxidase (SOX) was subsequently integrated onto the surface of spectrographic graphite rods for obtaining a glutathione biosensor. The working principle was based on the monitoring of O2 consumption that correlates the concentration of glutathione during the enzymatic reaction. A linear relationship between sensor response and concentration was obtained between 0.5 and 2.0 mM for oxidized glutathione (GSSG), and 0.2–1.0 mM for reduced glutathione (GSH) in the presence of 2 μM nicotinamide adenine dinucleotide phosphate (NADPH) under the optimum working conditions. Also, reduced/oxidized glutathione were separated by HPLC and utility of bienzymatic system was investigated as an electrochemical detector for the analysis of these compounds. All data were given as a comparison of two systems: biosensor and diode array detector (DAD).

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, , , , ,