Abstract
A CPE-HPLC (UV) method has been developed for the determination of Prometryne. In this method, non-ionic surfactant Triton X-114 was first used to extract and pre-concentrate Prometryne from water and soil samples. The separation and determination of Prometryne were then carried out in an HPLC-UV system with isocratic elution using a detector set at 254 nm wavelength. The parameters and variables that affected the extraction were also investigated and the optimal conditions were found to be 0.5% of Triton X-114 (w/v), 3% of NaCl (w/v) and heat-assisted at 50 °C for 30 min. Using these conditions, the recovery rates of Prometryne ranged from 92.84% to 99.23% in water and 85.48% to 93.67% in soil, respectively, with all the relative standard deviations less than 3.05%. Limit of detection (LOD) and limit of quantification (LOQ) were 3.5 μg Lâ1 and 11.0 μg Lâ1 in water and 4.0 μg kgâ1 and 13.0 μg kgâ1 in soil, respectively. Thus, we developed a method that has proven to be an efficient, green, rapid and inexpensive approach for extraction and determination of Prometryne from soil samples.
Related Topics
Physical Sciences and Engineering
Chemistry
Analytical Chemistry
Authors
Jihai Zhou, Jiandong Chen, Yanhong Cheng, Daming Li, Feng Hu, Huixin Li,