Article ID Journal Published Year Pages File Type
1244903 Talanta 2009 5 Pages PDF
Abstract

A new methodology for the in-line preconcentration, clean-up and speciation of mercury by use of an anion-exchange membrane is proposed. The speciation of mercury is based on retention of its tetrachloro complex onto the membrane while organic mercury flows freely through it. A multisyringe is used as a liquid driver and a cold vapour atomic fluorescence detector is employed to ensure a high sensitivity. Organic mercury is decomposed into to inorganic mercury by using a UV lamp. The carrier and reductant streams consist of 1.5% (m/v) hydrochloric acid and 2% (m/v) tin chloride, respectively. Certified reference material DORM-2 was digested with 37% hydrochloric acid and analysed directly without the need for extraction. The proposed system is more environmental friendly than the classical liquid–liquid extraction procedure. Mercury recoveries from spiked samples and the reference material were all close to 100%. An LOD of 14 and 16 ng/L was obtained for total and organic mercury, respectively, both with an RSD less than 1.3%.