Article ID Journal Published Year Pages File Type
1244914 Talanta 2009 5 Pages PDF
Abstract

Ion-pairing chromatography coupled with inductively coupled plasma mass spectrometry (ICP-MS) used for the speciation of phosphorus is limited as the mobile phase containing organic solvents changes in detection sensitivity and the carbon precipitates on torch and cones. To address this issue, anion-exchange chromatography with ICP-MS has been used for the speciation of glyphosate, phosphate and aminomethylphosphonic acid in soil extracts. The separation of the targets on a new column was achieved within 5 min using an eluent containing 20 mM NH4NO3 at pH 5.1. Furthermore, since the polyatomic ions such as 14N16O1H+ and 15N16O+ from a nitrogen-based ion-pairing reagent interfered with ICP-MS detection of 31P, an octopole reaction system was investigated to determine whether the polyatomic interferences could be reduced. The results show that addition of He to the cell can benefit analyses by reducing such interferences, but at the expense of reduced sensitivity. The detection limits in the range of 1.0–1.5 μg L−1 (expressed as P) was achieved when 50 μL was injected using He as the collusion gas.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry