Article ID Journal Published Year Pages File Type
1245110 Talanta 2009 6 Pages PDF
Abstract

A novel flow injection (FI) on-line displacement solid phase extraction preconcentration and/or separation method coupled with FAAS in order to minimize interference from other metals was developed for trace silver determination. The proposed method involved the on-line formation and subsequently pre-sorption of lead diethyldithiocarbamate (Pb-DDTC) into a column packed with PTFE-turnings. The preconcentration and/or separation of the Ag(I) took place through a displacement reaction between Ag(I) and Pb(II) of the pre-sorbed Pb-DDTC. Finally, the retained analyte was eluted with isobutyl methyl ketone (IBMK) and delivered directly to nebulizer for measuring. Interference from co-existing ions with lower DDTC complex stability in comparison with Pb-DDTC, was eliminated without need for any masking reagent. With 120 s of preconcentration time at a sample flow rate of 7.6 mL min−1, an enhancement factor of 110 and a detection limit (3 s) of 0.2 μg L−1 were obtained. The precision (RSD, n = 10) was 3.1% at the 10 μg L−1 level. The developed method was successfully applied to trace silver determination in a variety of environmental water samples and certified reference material.

Related Topics
Physical Sciences and Engineering Chemistry Analytical Chemistry
Authors
, ,