Article ID | Journal | Published Year | Pages | File Type |
---|---|---|---|---|
1245129 | Talanta | 2009 | 5 Pages |
Based on well-known silver mirror reaction the Ag film was formed on Au film modified by self-assembled monolayer (SAM) of 1,6-hexanedithiol (HDT). The sensitivity of the biosensor based on this Ag/Au film is enhanced compared to that based on Au film. When the surface plasmon resonance (SPR) biosensor based on this Ag/Au film was used to determine human IgG, the range of concentrations of human IgG that could be determined is 0.30–40.00 μg mL−1. The lowest concentration (0.30 μg mL−1) that could be detected was about 8 times lower than that obtained by the biosensor without modification by Ag film (2.50 μg mL−1), which demonstrated that the biosensor based on Ag/Au film could make the resonant wavelength move to longer wavelength following with the sensitivity enhancement of the SPR biosensor.